CRETEM AP MD · MULTI-DOSE POUCH PACKAGING
Multi-dose capacity.
Made for your mix.
Choose 300, 360 or 480 dedicated medication cassettes. Select A for cassette dispensing, or B to add one FSP and a 60-cell MDU for suitable non-cassette medicines.

- 300–480
- Cassette positions · three capacities
- 17inch
- Published touchscreen size
- 60pouches/min
- Maximum unit-dose packaging speed
- A / B
- B adds 1 FSP and a 60-cell MDU
CAPACITY AROUND YOUR FORMULARY
Choose the preparation path around your medicines.
Compare cassette capacity and preparation method together. AP MD offers three capacities in two configurations, with different dimensions and weights for A and B.
Choose your cassette capacity
300, 360 or 480 positions, with small, medium and large cassette mixes. Match the cassette count to your regular medication list.
Individual cassette-drawer access
The supplied AP MD diagram identifies individual drawer locks. Plan staff access, refilling and key procedures around the cabinet.
Choose A or B preparation
A has no FSP or MDU. B adds one FSP and a 60-cell MDU for suitable non-cassette medicines.
Cabinet dimensions describe the machine body. Door movement, refilling, ventilation, connections and service access need additional space.
PRODUCT FEATURES
See what makes the workflow work.
Explore AP MD cabinet access, B-series preparation, controls and packing. Shared component details are identified alongside AP MD reference images.
Choose 300, 360 or 480 cassette positions. Each capacity has a published small, medium and large cassette mix, allowing the configuration to be assessed against the medicines you regularly package.

Dedicated medication cassettesShared FS-series component detail The supplied AP MD diagram identifies individual drawer locks and single-drawer cassette storage. Review access, keys and local replenishment procedures with the proposed cabinet; locks do not establish regulatory compliance on their own.
The AP MD sheet specifies a 17-inch control panel rather than the AP FS series’ 10.1-inch screen. Use the agreed interface or internal order entry. The sheet lists Windows 10; confirm the supplied OS and support lifecycle.
AP-300MD and AP-360MD share an 896 mm main-body width. AP-480MD is 1,174 mm wide. All have a listed 1,070 mm depth; height and weight depend on the A/B version.
The supplied A-series table lists no FSP and no MDU. Choose this preparation arrangement only after reviewing your medication mix and the work that must be handled outside cassette dispensing.
B-series models add ONE Free Shape Packing unit and a 60-cell Manual Divide Unit. This is the AP MD table’s configuration; the AP FS series’ 1–2 FSP options should not be carried over.

FSP and MDU are different preparation paths. B-series Free Shape Packing supports suitable non-cassette medicines; its 60-cell Manual Divide Unit is prepared by staff. These modules are not listed on A-series models.
FROM ORDER TO POUCH
One connected packaging sequence.
Automatic dispensing and packaging sit within your team’s preparation, verification and release process.
Receive the order
Use the agreed OCS interface or internal entry on the 17-inch touchscreen.
Dispense from cassettes
Dedicated medication cassettes dispense through an infrared-monitored path.
Prepare the B-series additions
On B models, use the specified FSP and 60-cell MDU for suitable non-cassette medicines.
Print and seal
Apply the configured label information and seal each medication pouch.
Check and release
Complete finished-pouch content checks, exception handling and pharmacy release procedures.
Infrared dispensing detection supports the process. Finished-pouch content checks and final release remain with the pharmacy.
CHOOSE YOUR AP MD CONFIGURATION
Three capacities. Two preparation paths.
Choose cassette capacity, then select A or B. See how the preparation modules, dimensions and weight change.
AP-300MD-A
132 small · 120 medium · 48 large cassettes
- Main-body height
- 2,021 mm
- Published weight
- 650 kg
Compare the AP MD configurations
| Configuration | AP-300MD-A | AP-300MD-B | AP-360MD-A | AP-360MD-B | AP-480MD-A | AP-480MD-B |
|---|---|---|---|---|---|---|
| Medication cassettes | 300 | 300 | 360 | 360 | 480 | 480 |
| Cassette mix · small / medium / large | 132 / 120 / 48 | 132 / 120 / 48 | 156 / 144 / 60 | 156 / 144 / 60 | 208 / 192 / 80 | 208 / 192 / 80 |
| Free Shape Packing | No FSP | 1 FSP | No FSP | 1 FSP | No FSP | 1 FSP |
| Manual Divide Unit | No MDU | 60 cells | No MDU | 60 cells | No MDU | 60 cells |
| Main body · W × D × H (mm) | 896 × 1,070 × 2,021 | 896 × 1,070 × 1,907 | 896 × 1,070 × 2,021 | 896 × 1,070 × 1,907 | 1,174 × 1,070 × 2,141 | 1,174 × 1,070 × 2,027 |
| Main body · W × D × H (ft, approx.) | 2.94 × 3.51 × 6.63 | 2.94 × 3.51 × 6.26 | 2.94 × 3.51 × 6.63 | 2.94 × 3.51 × 6.26 | 3.85 × 3.51 × 7.02 | 3.85 × 3.51 × 6.65 |
| Published machine weight | 650 kg | 670 kg | 650 kg | 670 kg | 800 kg | 820 kg |
- Packaging speed
- Up to 60 unit-dose pouches/min; brochure average 55 multi-dose pouches/min
- Control panel
- 17-inch touchscreen; supplied brochure lists Windows 10. Confirm supplied OS and support lifecycle.
- Dispensing detection
- Infrared photodetection; finished-pouch verification remains required
- Medication storage
- Cabinet-type storage with cassettes in single drawers and individual drawer locks
- Pouch printing
- Thermal transfer; configurable printed content and layout
- Wrapping paper size
- 75 × 70 mm
- Order input
- OCS (Order Communication System) linkage or internal entry; confirm interface compatibility
- Listed power consumption
- 500 W; maximum 1,000 W
- Regional supply options
- 220 V / 60 Hz (Korea); 120 V / 60 Hz (US); 230–240 V / 50 Hz (Europe). Confirm the supplied electrical configuration.
- Manufacturer / origin
- CRETEM / Republic of Korea
Based on the supplied AP MD A/B specification sheet. A models list no FSP or MDU; B models list one FSP and a 60-cell MDU. Confirm the current model, modules, installed envelope, weight, electrical requirements and operating system in your MAXIS proposal. Packaging speed is not end-to-end prescription throughput.
PLAN THE COMPLETE INSTALLATION
A good fit goes beyond the cabinet.
Bring your floor plan, medicine list and pharmacy software details. Build the proposal around the equipment and the work around it.
Software and order flow
Review the OCS interface or internal entry, required prescription fields, network requirements and representative test orders. Confirm the supported pharmacy interface and version with MAXIS before planning automatic order transfer.
Access, training and care
Confirm floor loading and delivery access, installation clearances, cleaning between medicines, cassette replenishment and printer maintenance. Confirm operator training, warranty, spare parts and service terms in the quotation.
Materials and ongoing costs
Include the approved wrapping material, print ribbon and reorder requirements. Validate both pouch format and printed information with samples from the configuration proposed.
BEFORE YOU DECIDE
Your questions. Clearer answers.
The practical details of AP MD capacity, A/B preparation and installation.
What is the difference between A and B models?
In the supplied AP MD sheet, A configurations have cassette dispensing with no FSP and no MDU. B configurations add one FSP and a 60-cell MDU. Each cassette capacity is offered in both versions. Dimensions and weight differ, so select the actual version rather than treating B as an unspecified add-on.
Which cassette capacities are listed?
AP-300MD has 300 cassette positions (132 small, 120 medium, 48 large); AP-360MD has 360 (156/144/60); AP-480MD has 480 (208/192/80). These are medication positions, not a fixed number of patients.
Is AP MD the same as the AP Model?
No. AP MD is a separate cabinet range with individual drawer locks, a listed 17-inch control panel and A/B configurations. The AP FS range has different model capacities and preparation options. Use the relevant specification and installation drawing for each.
Can an A model be converted to B later?
The supplied sheet defines separate A/B configurations and different heights and weights. It does not establish a field-conversion path. Confirm any upgrade or retrofit directly with MAXIS before purchasing a configuration around that assumption.
Does B-series have one or two FSP units?
The AP MD table explicitly lists one FSP on each B model and none on A models. Generic references to two FSP units in other platform copy do not override this model-specific table. Confirm the proposed machine’s actual modules in the quotation.
What packaging speed is listed?
The supplied sheet lists up to 60 unit-dose pouches per minute and an average of 55 multi-dose pouches per minute. These are packaging figures, not end-to-end prescription output. Preparation, refilling, checking, cleaning and medication mix also affect throughput.
What pouch material size does the sheet specify?
The supplied AP MD sheet lists 75 × 70 mm wrapping and thermal-transfer printing. Confirm the compatible material, ribbon, text layout and barcode readability on actual pouch samples. No alternate wrapping size is assumed from another AP range.
Do detection and auto-repacking replace pouch checks?
No. Infrared photodetection monitors dispensing, and supported error recovery deals with recognised packing errors. Your pharmacy’s medication-content verification, exception handling and final release process remain required.
What space and weight should we plan for?
The 300/360 models have a published main-body footprint of 896 × 1,070 mm, and 480 models are 1,174 × 1,070 mm. Listed weights range from 650 to 820 kg. Allow additional working/service space and confirm floor loading, delivery access and installed dimensions for the selected A/B model.
Will AP MD connect to our pharmacy software?
The supplied sheet lists OCS linkage or internal entry. Provide your software name/version, data fields, network requirements and sample orders. MAXIS needs to confirm the interface, responsibilities and acceptance tests for the actual AP MD configuration.
What electrical and IT details need confirmation?
The sheet lists 500 W consumption, maximum 1,000 W, and regional supply options. It also lists Windows 10 with a 17-inch touchscreen. Confirm the supplied operating system/support lifecycle, electrical configuration, network and remote-support requirements before installation.
What support and ongoing supplies should be included?
Confirm training, approved cleaning, wrapping material, thermal-transfer ribbon, spare parts, warranty, service scope and recurring costs in your MAXIS proposal. Stock management is optional in the shared brochure, so establish whether it is included rather than assuming it is standard.
COMPARE YOUR OPTIONS
Different capacity. The same focus on fit.
YOUR FLOOR PLAN. YOUR MEDICATION MIX.
Let’s find your AP MD configuration.
Bring the space you have and the workflow you want to build.
We’ll help you assess the right configuration.